| Model Number |
8401 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had a low reading and not accurate, no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device's seal was intact.There was no physical damage on the device.The technician powered on off unit and a visual inspection of the flow rate test, carbon dioxide c02 test.The reported complaint was not verified after test the c02 function on the return monitor and the c02 reading was within specification.Ran c02 10% from a 4086 multi-parameter simulator.Low calibration was performed at 72 millimeters of mercury (mmhg).Flow rate was tested at 140 milliliter/minute (ml/min) (spec 120 plus /minus 20 ml/min.) the root cause of the reported issue could not be confirmed as the reported issue could not be replicated at the time of the device evaluation.The technician replaced internal tube kit.
|
| |
|