| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product has been received by zimmerbiomet.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the trial poly size g 10mm broke during removal.No pieces fell into the patient.No patient impact.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information. visual examination of the returned product identified signs of use (nicked/gouged) and the device is fractured into pieces.Review of the device history records identified no deviations or anomalies during manufacturing.Reported event is not related to a combination of products; therefore a compatibility review is not applicable.Medical records were not provided.Complaint is confirmed.The root cause of the reported issue is attributed to wear over the course of the 5+ years this device was in the field.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|