| Model Number |
DBP-150CLASS145 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A diamondback 360 peripheral orbital atherectomy device was used to treat the anterior tibial artery.Fluoroscopy imaging was performed during a treatment and it was observed that the crown was not moving with the tip of the driveshaft.The oad was removed, and the crown was observed to be moving up and down the shaft.A new oad was used to complete the procedure.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The oad was returned to csi without the guidewire.Analysis revealed the crown to be detached and located proximal to its original location on the driveshaft.Scanning electron microscopy analysis of the crown identification and driveshaft crown bond section revealed the presence of adequate solder residue indicating a sufficient crown bond present prior to the separation.Diagnostic analysis identified treatment runs averaging forty-five seconds.The instructions for use states: "a rest period of thirty seconds is recommended for every thirty seconds of treatment, with a maximum treatment of eight minutes per oad." the root cause of the crown detachment was unable to be determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
| |
|