| Model Number |
IPN028442 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Reported issue: defective drainage system with bubbling when it should not.The air leak was found at the re/blue connection.The device was changed.No clinical consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The device was returned for investigation.A visual exam was performed and it was observed that the red connector from the ats was not received.The device history record of lot number 74k2200798 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint.The complaint cannot be confirmed as an incomplete sample was returned.Both connectors are required to perform a proper investigation and determine a root cause.
|
| |
|
Event or Problem Description
|
|
Reported issue: defective drainage system with bubbling when it should not.The air leak was found at the re/blue connection.The device was changed.No clinical consequences.Additional information received on 14 april 2023 states that the patient's condition did not worsen after the event, there was no injury to the patient, and that there was no medical intervention.The leak occurred while transporting the patient back from surgery and it was an air leak that occurred on the red/blue connection.
|
| |
|