• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND PEDIATRIC AND NEONATAL TRACH TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA FLEXTEND PEDIATRIC AND NEONATAL TRACH TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 60PFSS45
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the item was "faulty, the coils on damaged from production".No patient involvement reported.
 
Additional Manufacturer Narrative
Date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Additional Manufacturer Narrative
One device was received for investigation.The reported issue was confirmed during visual inspection, which identified damage to the device tube.A review of manufacturing device history records for the reported lot number was conducted, and no discrepancies or anomalies were identified.Per the complaint description, the damage was not observed until its use in the patient.No damage was observed during the device pre-check.Based on the results of the inspection and the description of the event, the investigation determined the root cause could be due to improper use.No actions have been assigned at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA FLEXTEND PEDIATRIC AND NEONATAL TRACH TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16744072
Report Number3012307300-2023-04138
Device Sequence Number5378480
Product Code JOH
UDI-Device Identifier15021312516791
UDI-Public15021312516791
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number60PFSS45
Device Catalogue Number60PFSS45
Device Lot Number4347209
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2023
Supplement Date Received by Manufacturer06/29/2023
Initial Report FDA Received Date04/14/2023
Supplement Report FDA Received Date06/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-