Catalog Number 8065751763 |
Device Problem
Inability to Irrigate (1337)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A cutomer reported that during surgery an ophthalmic operating console fluidics was unavailable.Procedure was completed by restating the system.There was no patient harm.Additional information has been requested.
|
|
Manufacturer Narrative
|
The company representative was able to confirm but was unable to replicate the reported event.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The system was found to functionally exhibit no problems.Therefore, the root cause of the reported event is inconclusive.Manufacturers will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|