| Model Number |
M0066401091 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Detachment of Device or Device Component (2907); Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6 (device codes): device code a0501 captures the reportable event of core wire broken.Block h10: the returned amplatz super stiff was analyzed.Upon visual assessment, it was observed that the coil wire was unraveled, and the core wire was detached at 2.6 cm from distal tip section to the transition point.The reported event was confirmed.Based on the condition of the returned device, engineers determined that problems were traced to manufacturing process.Boston scientific has determined the most probable cause of this complaint is manufacturing deficiency.An investigation to address this problem is in progress.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an amplatz super stiff was used during a percutaneous nephrolithotomy procedure on (b)(6) 2023.During procedure, the guidewire had difficulty advancing with 8/10 dilator and got damaged like split, frayed and disheveled.The procedure was completed with a new amplatz super stiff.There were no patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation finding of the coil wire was unraveled, and the core wire was detached at distal tip section.
|
| |
|