• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE TAPERED CERTAIN IMPLANT 5 X 15MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE TAPERED CERTAIN IMPLANT 5 X 15MM; DENTAL IMPLANT Back to Search Results
Model Number XIFNT515
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 03/02/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth site 18 was removed due to peri-implantitis.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Patient weight unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text : summary investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE TAPERED CERTAIN IMPLANT 5 X 15MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16746686
Report Number0001038806-2023-00749
Device Sequence Number19649642
Product Code DZE
UDI-Device Identifier00844868022039
UDI-Public(01)00844868022039(17)210805(10)2016071440
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K130949
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/05/2021
Device Model NumberXIFNT515
Device Catalogue NumberXIFNT515
Device Lot Number2016071440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2023
Initial Report FDA Received Date04/14/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/05/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
Patient SexFemale
-
-