MAUDE Adverse Event Report: CARDINAL HEALTH CONNECT FEED STE 1000ML NS; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH CONNECT FEED STE 1000ML NS; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 77100FD
Medical Device Problem Codes Detachment of Device or Device Component (2907); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
Customer reports: i entered the patient's room and found the feeding tube leaking and with air in the tubing (feed was complete).There was approximately 20ml of formula on the ground.I looked closer at the set and noticed the connection between the tubing from the bag and the connection to the pump was not sealed and disconnected completely as i was inspecting it.This is at least the third instance of a kangaroo connect feeding set leaking like this that i have heard about.I notified the charge nurse, bagged the item, and left it on the desk.
 
Additional Manufacturer Narrative
A device history record review could not be reviewed since a valid lot number was not provided.One used sample was received at the manufacturing facility for evaluation.A visual and functional inspection was performed and found that one line detached from the cassette.The set was received broken.The upstream tube (the tubing before the purple cassette) and downstream tube (the tubing after the purple cassette) were not connected properly to the set.Both tubes seemed to be properly sized so that they would fit in the set under normal circumstances, but both tubes were disconnected from the set.This was checked by successfully placing both tubes in the proper places in the purple cassette to check if they fit.The disconnected upstream and downstream tubing both lacked the distinct ¿cloudy¿ look that tubes have when a solvent adhesive is present.This indicates that there was not enough solvent adhesive placed on the tube during the manufacturing process to withstand normal use.The root cause and corrective action investigation will be documented through a formal corrective/preventative action (capa).
 
Additional Manufacturer Narrative
Corrected data: section d4: lot number - the original lot number does not correspond to the reported product therefore the lot number has been updated to "unknown." this field should remain blank.Section g: contact office - manufacturing site - the reported lot number was reported incorrectly therefore the manufacturing site has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECT FEED STE 1000ML NS
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16746701
Report Number9612030-2023-03664
Device Sequence Number18281743
Product Code LZH
UDI-Device Identifier10884521190825
UDI-Public10884521190825
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number77100FD
Device Catalogue Number77100FD
Device Lot Number2227034764
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2023
Supplement Date Received by Manufacturer03/29/2023
03/29/2023
Initial Report FDA Received Date04/14/2023
Supplement Report FDA Received Date05/12/2023
08/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-