| Model Number |
60PFSS30 |
| Medical Device Problem Codes |
Fracture (1260); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Other, other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer had 2 incidents of split flanges on the tracheostomy tubes.Occurred while in use with patient, a trach tube change was required.No patient injury reported.(b)(4) documents that first product malfunction and (b)(4) documents the second product malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Email is: regulatory.Responses@icumed.Com.Two device samples were received in used condition without the original packaging.Per visual inspection, it was detected a cut in the flange/neck strap.The complaint was confirmed.No other analysis was performed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.The cause for the condition was due to manufacturing.The complaint fault has been escalated to address a full root cause investigation for damaged flange/neck strap issues and is currently in the investigation phase.
|
| |
|