| Catalog Number |
824M6545 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2023-00114.
|
| |
|
Event or Problem Description
|
|
It was reported that a closure top was found to have migrated out of the mating vital mis screw via x-ray taken at the patient's one month postoperative appointment.A revision surgery is scheduled to remove and remove and replace the screw and closure top.This is report two of two for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that a closure top was found to have migrated out of the mating vital mis screw via x-ray taken at the patient's one month postoperative appointment.A revision surgery was performed to remove and replace the screw and closure top.The revision surgery was a success without any issues.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d4: lot & udi number, d9, and h3.Additional information in b5, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection of the returned device revealed signs of use, but no signs of any damage.A functional inspection was performed with the returned screw (pn 824m6545 lot sbm141994) and found the closure top was able to mate with the screw and fully seat into the tulip head.Review of the provided x-ray images shows that the closure top has clearly migrated out of the screw.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|