• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 1-2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 1-2 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71453201
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Metal Related Pathology (4530)
Date of Event 02/02/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).Cañizares, s., barriga, g.C.C., jarrín, f.V., freire, c.D.P., & freire, c.D.P.(2023).Metallosis after oxinium total knee arthroplasty in a patient with rheumatoid arthritis: a case report.Cureus journal of medical science, 15(2).This complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from literature sources that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that on literature review "metallosis after oxinium total knee arthroplasty in a patient with rheumatoid arthritis: a case report", one (1) patient presented swelling and a growing mass in the internal angle of the knee after an index tka procedure using a genesis ii system.This adverse event was treated via revision surgery.The post of the polyethylene liner was detached, the liner dislocated towards the internal anterior knee compartment and delamination of the oxinium femoral component was noted.Additionally, a profuse metallosis was observed.A synovectomy was performed, and the insert was replaced with a new one of the same size.The patient was discharged after one week.
 
Additional Manufacturer Narrative
D1, d2a, d4 (device information), d10 (concimitant devices), g4 (510k), h4 (manufaturing date).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 1-2 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16747001
Report Number1020279-2023-00836
Device Sequence Number8576254
Product Code JWH
UDI-Device Identifier00885556033784
UDI-Public00885556033784
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Literature
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/13/2025
Device Model Number71453201
Device Catalogue Number71453201
Device Lot Number15LT61553
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2023
Supplement Date Received by Manufacturer04/14/2023
Initial Report FDA Received Date04/14/2023
Supplement Report FDA Received Date04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PN: 71420162 / LOT: 15JM19941; PN: 71421014 / LOT: 16AM00924; UNKN GENESIS II TOTAL KNEE FEM COMP OX CEM
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-