• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIPATH IMAGING, INC BD PREPSTAIN¿; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIPATH IMAGING, INC BD PREPSTAIN¿; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED Back to Search Results
Model Number 490100
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter phone code: (b)(6).H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd prepstain¿ hydraulic rods do not hold up.No injuries have been reported.The following information was provided by the initial reporter: tripath us ii - centrifuge hydraulic rods won't hold up.
 
Additional Manufacturer Narrative
H6.Complaint reports lid failure on centrifuge associated with prepstain (catalog number 490100) serial number (b)(6).Complaint alleges centrifuge hydraulic rod failure, no support for the lid to stay open.Customer was informed to replace hydraulic rods or recommend replacing the centrifuge.Customer was advised that the centrifuge is out of warranty.Customer advised no one was hurt and issue was not resolved.Root cause is not determined, and this complaint is not a confirmed failure of the instrument based on the customer response.Dhr review is not required because the part that allegedly failed is not tested as part of the manufacturing process but is shipped separately and tested during installation.Review of risk management files confirms there are no new or modified risks associated with this failure mode.There are no corrective action plans or other corrections occurring.Bd quality will continue to monitor for trends associated with failure of "centrifuge." h3 other text : see h10.
 
Event or Problem Description
It was reported that bd prepstain¿ hydraulic rods do not hold up.No injuries have been reported.The following information was provided by the initial reporter: tripath us ii - centrifuge hydraulic rods won't hold up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PREPSTAIN¿
Common Device Name
PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED
Manufacturer (Section D)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer (Section G)
TRIPATH IMAGING, INC
1022 corporate park drive
mebane NC 27302
Manufacturer Contact
phillip emmert
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key16747802
Report Number3008007472-2023-00002
Device Sequence Number11620289
Product Code MKQ
UDI-Device Identifier00382904901004
UDI-Public00382904901004
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P970018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number490100
Device Catalogue Number490100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2023
Supplement Date Received by Manufacturer05/12/2023
Initial Report FDA Received Date04/14/2023
Supplement Report FDA Received Date05/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-