C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA 4F DL MBPLUS (3CG, GUARDIVA, PROBE COVER); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Catalog Number S1274108D2
Medical Device Problem Codes Fluid/Blood Leak (1250); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer that multiple same lot numbers occluding within short time after placement.Using additional kits because of leaking lumens.No other information was provided.The customer reported multiple events however the exact number of events was not provided to bd vascular access devices field assurance.
 
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.
 
Event or Problem Description
It was reported by the customer that multiple same lot numbers occluding within short time after placement.Using additional kits because of leaking lumens.No other information was provided.The customer reported multiple events however the exact number of events was not provided to bd vascular access devices field assurance.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of leaking lumens for a powerpicc is unconfirmed because the problem could not be reproduced.One 4 fr d/l powerpicc was returned for evaluation.An initial visual observation showed use residues throughout the returned device, and nothing remarkable was observed during initial visual observations.A functional test of infusing and pressurizing both lumens with water using a 12 ml syringe were successful, and no leaking was observed from either lumen.Both lumens were patent to infusion without any resistance.Because the device failure could not be reproduced, the complaint of leaking powerpicc lumens is unconfirmed.Possible contributing factors may include placement of the device in the patient¿s vasculature, causing an occlusion in the device.H3 other text : evaluation findings are in section h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA 4F DL MBPLUS (3CG, GUARDIVA, PROBE COVER)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
suedabba mahboobi
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key16748278
Report Number3006260740-2023-01354
Device Sequence Number3429337
Product Code LJS
UDI-Device Identifier00801741154843
UDI-Public(01)00801741154843
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K180548
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Catalogue NumberS1274108D2
Device Lot NumberREGZ0089
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/23/2023
Supplement Date Received by Manufacturer06/26/2023
Initial Report FDA Received Date04/14/2023
Supplement Report FDA Received Date07/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient SexFemale
-
-