| Model Number |
VENUM16100 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 08/2024).
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure via internal jugular access, stent allegedly adhered to the delivery system for approximately twenty five to thirty seconds and then released.It was further reported that the health care provider allegedly required to slightly twist the delivery system to get the stent to fully release.Reportedly the stent was successfully deployed and covered the lesion.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not available for evaluation.One provided image of an x-ray screen demonstrates the deployed stent with lying wire inside the vessel; the resolution is poor so that single struts are not visible in detail.Furtherly, a single image does not indicate a time period, so that a clear statement regarding the stent holding process by the inner catheter is not possible.Based on the information available the investigation is closed with inconclusive result for stent expansion issue.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential worst case complications and adverse events potentially related to expansion issue such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, stent fracture, and malposition.H10: d4 (expiry date: 08/2024), g3.H11: h6 (method, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure via internal jugular access, stent allegedly adhered to the delivery system for approximately twenty five to thirty seconds and then released.It was further reported that the health care provider allegedly required to slightly twist the delivery system to get the stent to fully release.Reportedly the stent was successfully deployed and covered the lesion.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|