• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SYNPOR SQUARE SHEET-STERILE 50MMX50MM/0.8MM THICK; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SYNPOR SQUARE SHEET-STERILE 50MMX50MM/0.8MM THICK; MESH, SURGICAL Back to Search Results
Model Number 08.510.120S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: ftl, gwo.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: part:08.510.120s.Lot:ds7007483.Manufacturing site: werk selzach.Supplier:dsm biomedical.Release to warehouse date:01 jun 2021.Expiration date: 01 may 2026.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the synpor square sheet-sile 50x50/0.8 thick was found to be delaminated across several sections of the sheet and there are small flakes at the surface of the table.One of the fragments appears to be broken across the mid section.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the synpor square sheet-sile 50x50/0.8 thick.Relevant actions have been taken to address the issue.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in china as follows: it was reported that during a rhinoplasty surgery on (b)(6) 2023, during cutting, it was noted that the device was broken off.Another device was used to complete the surgery, and there were no adverse consequences to the patient.There was no delay to surgery.No fragments were retained in the patient.This report involves one synpor square sheet-sterile 50mmx50mm/0.8mm thick.This is report 1 of 1 for pc-(b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the synpor square sheet-sile 50x50/0.8 thick was found broken in four parts, and was found to be delaminated in some sections of the product, one of the fragments appears to be broken across the mid section.No other issues were found.A dimensional inspection was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the synpor sheet 50*50 t0.8 would contribute to the complained device issue.Based on the investigation findings, the root cause is traced to design.The product issue has been addressed through depuy synthes quality system.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNPOR SQUARE SHEET-STERILE 50MMX50MM/0.8MM THICK
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK SELZACH
bohackerweg 5
selzach 2425
SZ   2425
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16750371
Report Number8030965-2023-04698
Device Sequence Number11620194
Product Code FTM
UDI-Device Identifier10887587026039
UDI-Public(01)10887587026039
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051879
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2026
Device Model Number08.510.120S
Device Catalogue Number08.510.120S
Device Lot NumberDS7007483
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/24/2023
Supplement Date Received by Manufacturer04/24/2023
06/12/2023
Initial Report FDA Received Date04/17/2023
Supplement Report FDA Received Date05/09/2023
06/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient Age26 YR
Patient SexFemale
Patient Weight49 KG
-
-