Investigation results: returned qty.( b)(4) inserts, 1.82005, 22291-00085016, gd, bul, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: meets spec, does not meet spec: n/a, tip condition: meets spec, does not meet spec: n/a, stack condition: meets spec, does not meet spec: n/a.Tested on: g136 gage id#: 9626-2, due date: 03/31/2023, set pressure: 35 psi, water flow: output rate: 35 psi meets spec, does not meet spec: n/a, spray pattern: meets spec, does not meet spec: n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a.Vibration: meets spec, does not meet spec: n/a.Grip condition: meets spec , does not meet spec: n/a.Other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#: 6116a, due date: 09/30/2023.The temperature was 87.0° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed / not confirmed, 2.82005, 22291-00085016, gd, bul, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: meets spec, does not meet spec: n/a.Tip condition: meets spec, does not meet spec: n/a, stack condition: meets spec, does not meet spec: n/a, tested on: g136 gage id#: 9626-2, due date: 03/31/2023 , set pressure: 35 psi, water flow: output rate: 35 psi meets spec, does not meet spec: n/a, spray pattern: meets spec does not meet spec n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a, vibration: meets spec, does not meet spec: n/a, grip condition: meets spec does not meet spec n/a other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#: 6116a, due date: 09/30/2023.The temperature was 90.1° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed / not confirmed.3.82004, 22300-00085643, 43, warranty, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: 3.12, meets spec, does not meet spec: n/a, tip condition: meets spec, does not meet spec: n/a, stack condition: meets spec, does not meet spec: n/a, tested on: g136 gage id#: 9626-2, due date: 03/31/2023, set pressure: 35 psi, water flow: output rate: 35 psi meets spec, does not meet spec: n/a, spray pattern: meets spec, does not meet spec: n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a, vibration: meets spec , does not meet spec: n/a, grip condition: meets spec, does not meet spec: n/a, other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#: 6116a, due date: 09/30/2023.The temperature was 86.8° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed / not confirmed, 4.82004, 22300-00085643, 43, warranty, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: 3.15, meets spec does not meet spec n/a tip condition: meets spec, does not meet spec: n/a, stack condition: meets spec , does not meet spec: n/a, tested on: g136 gage id#: 9626-2, due date: 03/31/2023, set pressure: 35 psi, water flow: output rate: 35 psi meets spec, does not meet spec: n/a, spray pattern: meets spec, does not meet spec: n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a, vibration: meets spec , does not meet spec: n/a, grip condition: meets spec, does not meet spec: n/a, other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#: 6116a, due date: 09/30/2023.The temperature was 88.9° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed / not confirmed.5.82004, 22305-00086045, 44, warranty, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: meets spec, does not meet spec: n/a, tip condition: meets spec, does not meet spec: n/a, stack condition: meets spec, does not meet spec: n/a, tested on: g136 gage id#: 9626-2, due date: 03/31/2023, set pressure: 35 psi, water flow: output rate: 35 psi meets spec, does not meet spec: n/a, spray pattern: meets spec, does not meet spec: n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a, vibration: meets spec, does not meet spec: n/a.Grip condition: meets spec, does not meet spec: n/a.Other visual observation: returned a new unopen packaged insert.Alleged event: confirmed / not confirmed.
|