Model Number 0673
Device Problems Connection Problem (2900); Device Dislodged or Dislocated (2923); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2023
Event Type  Injury  
Event Description
It was reported that the impedance associated with the right ventricular lead was abnormal and the lead was discovered to be detached from the implantable cardioverter defibrillator (icd).The lead was successfully replaced, and no additional adverse patient effects were reported.The lead is expected to be returned for analysis.
 
Event Description
It was reported that the impedance associated with the right ventricular lead was abnormal and the lead was discovered to be detached from the implantable cardioverter defibrillator (icd).The lead was successfully replaced, and no additional adverse patient effects were reported.The lead was returned for analysis.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.A stylet insertion test was also performed and indicated no blockage in the lead lumen.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646
*   00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16752753
MDR Text Key313424417
Report Number2124215-2023-17974
Device Sequence Number1
Product Code LWS
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P910073/S145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/15/2023
Device Model Number0673
Device Catalogue Number0673
Device Lot Number153691
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexFemale
-
-