Model Number 1156T/86
Device Problem Therapy Delivered to Incorrect Body Area (1508)
Patient Problem Discomfort (2330)
Event Date 03/30/2023
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2023-17741 / 2017865-2023-17743.It was reported that the patient's right ventricular high voltage lead delivered an inappropriate shock.The shock was due to over-sensed noise from the right ventricular sensing lead.High capture threshold was also noted.The patient presented for a lead revision.Prior to the procedure, the patient had experienced diaphragmatic stimulation from the left ventricular lead.The leads were explanted and replaced.The patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16753354
MDR Text Key313430308
Report Number2017865-2023-17747
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00541473450202
UDI-Public0541473450202
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2011
Device Model Number1156T/86
Device Lot Number0002662234
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/30/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/14/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FORTIFY ASSURA
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexMale
-
-