H10: h3, h6: the reported device was received for evaluation.There was no relationship found between the device and the reported event.A visual evaluation of the device showed the t1, t2, and suture are still on the needle.The t1 is at the tip of the needle and the t2 has been pushed forward onto the dimple behind the t1.The variable depth straw has been trimmed.A functional evaluation of the device, using a simulation meniscus, showed the t1 was deployed and implanted as intended.The actuator moved the t2 into position, was deployed, and implanted as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|