• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. SAW BLADES LINVATEC®/HALL® POWER SYSTEMS; SURGICAL SAW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. SAW BLADES LINVATEC®/HALL® POWER SYSTEMS; SURGICAL SAW Back to Search Results
Model Number N/A
Medical Device Problem Codes Break (1069); Loose or Intermittent Connection (1371); Detachment of Device or Device Component (2907); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the surgeon was preparing the tibia, the cementless keel saw blade snapped off in the bone.The hall power tool reciprocating attachment loosened from the hand piece and the saw blade twisted out.The event occurred during surgery.There were no adverse effects for patient apart from time wasted.There was a five-minute delay to get another pack of blades to complete the procedure.No additional medical intervention was required.Due diligence is complete and there is no further information available.
 
Additional Manufacturer Narrative
An investigation into the reported event has been initiated under cmp(b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.Initial reporter: telephone: (b)(6).Report source: foreign: new zealand.
 
Event or Problem Description
There is no additional information available.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Visual examination of the provided picture identified that the cementless keel saw blade head had broken off.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.H3 other text : product not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAW BLADES LINVATEC®/HALL® POWER SYSTEMS
Common Device Name
SURGICAL SAW
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16753732
Report Number0001526350-2023-00379
Device Sequence Number3366009
Product Code GFA
UDI-Device Identifier00889024494046
UDI-Public(01)00889024494046(17)280111(10)415767
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number11-3586
Device Lot Number415767
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2023
Supplement Date Received by Manufacturer10/23/2023
Initial Report FDA Received Date04/17/2023
Supplement Report FDA Received Date10/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
-
-