• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH STOPPER FOR LISS INSERTION GUIDES; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH STOPPER FOR LISS INSERTION GUIDES; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 324.019
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from france reports an event as follows: it was reported that upon receipt of the kit on february 8, 2023, when controlling the ancillary liss, the customer noticed that the piece 324.019 was broken.Also, the cover did not hold closed, so the screws were out of the rack.This report is for a stopper for liss insertion guides.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that stopper f/liss insert-guide was broken from one tab.The broken fragment was returned for examination.No other issues were found.A dimensional inspection for the stopper f/liss insert-guide was not performed as is not applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the stopper f/liss insert-guide would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A review of the device history records is not possible, the dhr is no longer available due to the age of the instrument (over 15 years old).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: g1 h4, h6: product code: 324.019.Lot: 9021278.Release to warehouse date : 01 january 2012.Expiration date : na.Supplier: na.Manufacturing site: werk bettlach.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STOPPER FOR LISS INSERTION GUIDES
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16755653
Report Number8030965-2023-04768
Device Sequence Number8556708
Product Code FZX
UDI-Device Identifier10886982190932
UDI-Public(01)10886982190932
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number324.019
Device Catalogue Number324.019
Device Lot Number9021278
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/27/2023
Supplement Date Received by Manufacturer05/09/2023
05/10/2023
Initial Report FDA Received Date04/17/2023
Supplement Report FDA Received Date05/09/2023
05/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: TRAUMA
Patient SexUnknown
-
-