| Model Number |
333851 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number 400595646.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: detail: needle broke off during spinal procedure for c section.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample was returned for evaluation.Visual evaluation of the used needle showed the tip to be broken/sheared off.Although the tip of the needle from the returned sample was broken off, since it was not broken or damaged upon the opening of the kit and it had been used in a procedure, per the manufacturer's investigation this defect is not likely to have happened during the manufacturing process.It is believed to happen during application.User will be referred to the ifu which states, "do not apply excessive force during needle advancement.Do not utilize needle if it bends or tip becomes damaged.If resistance is feltduring advancement of the needle, carefully correct the orientation of the needle but never apply excessive force." review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.
|
| |
|
Search Alerts/Recalls
|