| Model Number |
MI1200 SYNCHRONY PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma to the head.The user has no hearing sensation with the device anymore.Additional in situ measurements have been scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device was affected after a trauma to the head.The user had no hearing sensation with the device anymore.The user was reimplanted on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: the investigation results revealed overload fractures in the active electrode lead confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|