| Model Number |
SRL-140 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/29/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture, baker grade iv" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iv.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right side capsular contracture, baker grade iv.Device explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe, since it is a medical event.And is not related to the device.Additional observations: one opening on anterior side assessed, as surgical damage.White particles were observed, on the shell.
|
| |
|
Search Alerts/Recalls
|