C.R. BARD INC. (COVINGTON) -1018233 BARD® EXTENSION CABLE Back to Search Results
Model Number 153622
Medical Device Problem Codes Fitting Problem (2183); Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the temperature display was incorrect while using the extension cord for temperature sensing foley catheter.There was poor meshing of catheter and cable.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event is unconfirmed as the reported failure could not be reproduced.Photo samples were submitted, however, could not be evaluated for the reported failure.Two extension cables were returned without the original packaging.The cables were labeled 1 - 2 for identification purposes.No damage to the cord was noted.Connections of the cords to the plug and socket were secure.Using a multimeter, the cords were tested for and found to have continuity.The cords were connected to an in-house temperature sensing catheter submerged in a water bath (119118 lot ngevz327) set at 37 degrees celsius, both cords connected to the catheter snuggly with a click and required force to be removed.The cords were then attached to a kilo device, both connected snuggly and required force to be removed.The temperature displayed 37.2 degrees celsius, for both cables, which is similar to the thermometer readings.The samples meet the specification "plug and jack must be firmly secured to wire.As the reported event is unconfirmed, a dhr review is not required.As the reported event is unconfirmed, a label/packaging review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the temperature display was incorrect while using the extension cord for temperature sensing foley catheter.There was poor meshing of catheter and cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® EXTENSION CABLE
Common Device Name
EXTENSION CABLE
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16760643
Report Number1018233-2023-02610
Device Sequence Number3061877
Product Code EZL
UDI-Device Identifier00801741045875
UDI-Public(01)00801741045875
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number153622
Device Catalogue Number153622
Device Lot NumberBMGT0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2023
Supplement Date Received by Manufacturer07/13/2023
Initial Report FDA Received Date04/18/2023
Supplement Report FDA Received Date07/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;