MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problems
Migration or Expulsion of Device (1395); Obstruction of Flow (2423); Infusion or Flow Problem (2964)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of d10: product id 8780 lot# serial# (b)(4) implanted: (b)(6) 2022 explanted: (b)(6) 2023 product type catheter section d information references the main component of the system.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 08-sep-2024, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider via a company representative regarding a patient receiving fentanyl 632.5mcg/ml at 126.50mcg/day, clonidine 253mcg/ml at 50.60mcg/day and bupivacaine 12.7mg/ml at 2.54mg.Day via an implantable pump.It was reported the patient had a rebound of symptoms and the healthcare provider was unable to aspirate catheter during last refill appointment.The environmental, external or patient factors that may have led or contributed to the issue was noted as ¿n/a¿.Surgical intervention uncovered a significantly ¿kinked¿ catheter due to a tremendous amount of ¿flipping¿.The old catheter was removed and replaced.The issue was resolved at the time of the report, and it was noted that the healthcare provider would not have any further information regarding the event.
|
|
Search Alerts/Recalls
|
|
|