DEPUY IRELAND - 3015516266 PINNACLE DM LINER 56_49; TOTAL HIP JOINT REPLACEMENT PROSTHESIS Back to Search Results
Model Number 1218-56-049
Medical Device Problem Codes Off-Label Use (1494); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/21/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Dislocated pdm liner replaced with constrained liner.Doi: (b)(6) 2023.Dor: (b)(6) 2023.Unknown side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected:.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, ¿dislocated pdm liner replaced with constrained liner¿ the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the device found no damage or defects that could have contributed to the reported condition.Based on the observations, it is not possible to confirm a dislocation between the bi mentum altrx lnr 49 28 and the pinnacle dm liner 56_49.Additionally, the device was used for an indication not in the approved labeling, a dual mobility liner is not intended to be used with a bi mentum liner since these implants are from different systems and both are acetabular liners the overall complaint was confirmed for pinnacle dm liner 56_49.Based on the investigation findings, the potential cause can be traced to user due to off label use it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device [121856049/9828188] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE DM LINER 56_49
Common Device Name
TOTAL HIP JOINT REPLACEMENT PROSTHESIS
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16760911
Report Number1818910-2023-08126
Device Sequence Number11522469
Product Code LPH
UDI-Device Identifier10603295520290
UDI-Public10603295520290
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K200854
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-56-049
Device Catalogue Number121856049
Device Lot Number9828188
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Supplement Date Received by Manufacturer05/09/2023
12/07/2023
Initial Report FDA Received Date04/18/2023
Supplement Report FDA Received Date05/10/2023
12/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BI MENTUM ALTRX LNR 49 28; DLT TS CER HD 12/14 28MM +5.0; PINNACLE DM LINER 56_49; UNKNOWN HIP ACETABULAR CUP
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-