• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 4.0MM X 140MM, C-MOUNT, SPEEDLOCK, A/C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 4.0MM X 140MM, C-MOUNT, SPEEDLOCK, A/C Back to Search Results
Model Number 0502904030
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record, the reported failure "blurry" was not confirmed.According to henke: scope evaluated as a level 3 distal tip and fiber damaged.Probable root cause for this failure is: the complaint for ¿blurry¿ has not been confirmed and is due to customer used and handling.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Event or Problem Description
It was reported that there was blurry image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, PRECISION HD 30 DEG., ARTHROSCOPE, 4.0MM X 140MM, C-MOUNT, SPEEDLOCK, A/C
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16761978
Report Number0002936485-2023-00316
Device Sequence Number20331606
Product Code HRX
UDI-Device Identifier07613327063844
UDI-Public07613327063844
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K093677
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502904030
Device Catalogue Number0502904030
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2023
Supplement Date Received by Manufacturer03/22/2023
Initial Report FDA Received Date04/18/2023
Supplement Report FDA Received Date10/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-