Catalog Number 03.647.971
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: e3: reporter is a synthes employee.H3, h6: a device history record (dhr) review was not conducted, as manufacturing records related to the provided lot number were not available.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed, that screw removal screwdriver was twisted from the tip.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.After a visual inspection per guidance, it was determined, that the reusable instrument device was twisted from repeated use and servicing.Therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed, as the observed condition of screw removal screwdriver would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined, that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from portugal reports an event as follows: it was reported, that on (b)(6) 2023, the set only had one screwdriver shaft.One was missing.It was later determined, that the missing screwdriver shaft was present, but was stuck in the removal screwdriver.Upon manufacturer investigation, it was determined, that the removal screwdriver was twisted from the tip.This report is for a removal screwdriver.This is report 2 of 2 (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMOVAL SCREWDRIVER
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, CERVICAL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16763348
Report Number8030965-2023-04823
Device Sequence Number12307492
Product Code OVE
UDI-Device Identifier07611819404472
UDI-Public(01)07611819404472
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K112068
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.647.971
Device Lot Number3811P987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/17/2023
Initial Report FDA Received Date04/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCRDRIVER SHAFT T8 SELF-HOLDING; SCRDRIVER SHAFT T8 SELF-HOLDING
Patient SexUnknown
-
-