| Model Number |
IPN039296 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer called to report that the "belly bags she purchased online recently appear to have changed from what she has received in the 10+ years of use.She notes that the drain valve is much more difficult to twist and that the drain tube also seems to be a little longer".No report of patient harm or injury.Patient status is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported that "the user could not load a clip to the applier during use.Therefore, he/she replaced the applier with a new one to complete the procedure.Nothing fell/remained in the patient.On checking the applier later, the jaws were misaligned.The applier was sent to our technical specialist, who found that the handle of the applier was detached.".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device was returned and sent to the manufacturer (tecomet) for investigation.Tecomet reports: "the dhr for the returned instrument was reviewed and found complete without any irregularities.This instrument was produced at the tecomet, inc.Kenosha, wi facility as part of a(b)(4).Lot in march of 2022.It was found that this order was made from the correct materials and components and all approved processes were followed.Evaluation of the returned instrument shows that the jaws are slightly loose and misaligned to each other in the closed position.We are able to validate the complaint for the returned instrument.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) bosses are both damaged where they engage the jaws.We suspect that the damaged drive rod bosses caused the jaws to become slightly loose and misaligned in the closed position.We are unable to determine what caused the drive rod bosses to become damaged but mishandling of this device at the end user's facility is suspected.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings , no further actions will be taken in response to this complaint and this record will be deemed closed." other remarks: n/a corrected data: n/a.
|
| |
|