EXACTECH, INC. CR TIBIAL INSERT SZ 3, 9MM, SLOPE +; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
|
Back to Search Results |
|
| Model Number |
CR TIBIAL INSERT SZ 3, 9MM, SLOPE + |
| Medical Device Problem Codes |
Fracture (1260); Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Balance Problems (4401); Joint Laxity (4526); Metal Related Pathology (4530); Implant Pain (4561); Swelling/ Edema (4577)
|
| Date of Event |
06/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, fracture, instability, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|