It was reported by a healthcare professional in switzerland that during an unknown arthroscopic procedure on (b)(6) 2023, it was observed that the 4 mm barrel bur plus device stopped working and rotating.During in-house engineering evaluation, it was determined that the inner shaft on the device was seized; and that it was not possible to spin to any direction or separate.It was unknown if and how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).Investigation summary the complaint device was received for evaluation.Upon reviewing the device, it could be observed that the inner shaft is seized, it was not possible to spin to any direction or separate.No anomalies were found on the black connector.A manufacturing record evaluation was performed for the finished device m1811023 number, and no non-conformance were identified.As part of depuy synthese mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.A possible root cause for the jammed issue can be attributed to the procedural variables, such handling of the device or product interaction during procedure; higher speeds applied on the handpiece can increase the temperature of the device's shaft, lack of adequate irrigation and excessive side loading can contribute to a seized inner shaft, as per ifu: adequate irrigation to the tip of the blade or burr is required to cool the blade and prevent accumulation of excised material in the surgical site.Excessive side loading may result in blade wear and degradation, as well as clogging and/or seizing.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthese mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|