| Medical Device Problem Codes |
Computer Software Problem (1112); Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer alleged that the machine froze 99.4% into a scan and the gantry panels showed "radiating" with a patient in the gantry.Currently, there is no known injury to the patient.Issue is currently under investigation.
|
| |
|
Event or Problem Description
|
|
Gantry panels show "radiating" with patient in the gantry.
|
| |
|
Additional Manufacturer Narrative
|
|
"customer alleged that the machine froze 99.4% into a scan and the gantry panels showed ""radiating"" with a patient in the gantry.There are no known injuries.The investigation found that when the issue occurs, the scs appears to hang immediately after kv delivery completes.The couch continues to move into the bore until the user intervenes or the couch reaches the end stop.Possible causes include a deadlock or an unhandled exception that causes several threads to go down.In all instances, kv radiation stops at the expected time (before the issue arises).An anomaly was identified.Total risk is acceptable.The customer can clear the issue with a reboot.In conclusion, there have been no injuries.The system has been rebooted and no further actions are required.The issue is being tracked and fixed through the anomaly.".
|
| |
|