| Model Number |
60-05-60 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland has received a report of a scp electrical remote-controlled tubing clamp 500mm not working during a procedure.There is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in united kingdom.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected part was returned to manufacturer's site for a detailed investigation: the unit was disinfected, opened, cleaned, checked and tested.A loud noise was heard during erc opening and closing.To fix the issue, the motor area of the erc was cleaned, lubricated and greased.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2017.Based on investigation results and similar complaints, the reported event could be due to: - residues of dried fluids on the shafts, which obstruct the movements, generating noise; - wear of the unit ball bearing.
|
| |
|