| Model Number |
CD3231-40 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pocket Erosion (2013)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The sterilization records were reviewed and no evidence of abnormal sterilization cycle was found.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2023-17927, related manufacturer reference number : 2017865-2023-17928, related manufacturer reference number: 2017865-2023-17929.It was reported that the patient presented to hospital with infection and skin erosion at device pocket.The device pocket site was noted to be red since november 2022 and ruptured on (b)(6) 2023.The left ventricular lead was noted with high capture threshold.No intervention was performed.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
|
| |
|
Event or Problem Description
|
|
New information received notes the device, right atrial lead, right ventricular lead and left ventricular lead were explanted.The patient was in stable condition.
|
| |
|