BARD PERIPHERAL VASCULAR, INC. MISSION KIT; BIOPSY INSTRUMENT Back to Search Results
Model Number 2016MSK
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd-santo domingo.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 expiration date: 10/2024.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Event or Problem Description
It was reported that during a biopsy procedure, the needle allegedly broken into two pieces in the patient's lumbar disc.It was further reported that the patient had to be taken from ct to fluro to have this removed with the help of a trauma surgeon to remove the pieces.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a computed tomography guided lumbar disc biopsy procedure through normal density tissue, the needle allegedly broken into two pieces in the patient's lumbar disc.It was further reported that the inner stylet allegedly came out with the outer portion of the coaxial introducer.Furthermore, patient had to be taken from computed tomography to fluro with the help of a trauma surgeon to remove the pieces.Coaxial was used.Reportedly, the broken needle was removed by surgical cut.The patient is reported to be stable.
 
Additional Manufacturer Narrative
H10: additional information was received, and the file was reassessed for reportability and determined to be reportable as serious injury.H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd-santo domingo.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 10/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The manufacturing location was selected as unknown due to system limitations.The manufacturing location for the mission product is bd-santo domingo.Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were provided for review.First photo shows the needle inside the chamber where the needle was noted to be broken and it was noted to have residue.Second photo shows the trocar stylet inside the chamber where the needle was noted to be broken.Therefore, based on the photo review, the reported failure detachment can be confirmed.A definitive root cause for the alleged detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiration date: 10/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a computed tomography guided lumbar disc biopsy procedure through normal density tissue, the needle allegedly broken into two pieces in the patient's lumbar disc.It was further reported that the inner stylet allegedly came out with the outer portion of the coaxial introducer.Furthermore, patient had to be taken from computed tomography to fluro with the help of a trauma surgeon to remove the pieces.Coaxial was used.Reportedly, the broken needle was removed by surgical cut.The patient is reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MISSION KIT
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16768310
Report Number2020394-2023-00263
Device Sequence Number6417558
Product Code KNW
UDI-Device Identifier00801741097096
UDI-Public(01)00801741097096
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K171953
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2016MSK
Device Catalogue Number2016MSK
Device Lot Number0001442312
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/23/2023
Supplement Date Received by Manufacturer04/25/2023
08/18/2023
Initial Report FDA Received Date04/19/2023
Supplement Report FDA Received Date05/23/2023
08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Patient Weight91 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-