Model Number BOES505
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 12/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Summary investigation.
 
Event or Problem Description
It was reported that the doctor reports that the dental implant located in position number #4 failed due to peri-implantitis.The doctor reports that the procedure was not concluded by placing another implant.
 
Event or Problem Description
The doctor reports that the dental implant located in position number #4 failed due to peri-implantitis.The doctor reports that the procedure was not concluded by placing another implant.
 
Additional Manufacturer Narrative
Peri-implantitis is an infectious disease that causes inflammation of the gum and the bone structure around a dental implant.Chronic inflammation causes bone loss, which can lead to implant failure thus its removal.Investigations performed for hundreds of events where either of these conditions, infection, or bone loss, or both, have been reported, have concluded that the likely cause(s) for the implant failure in relation to these conditions are external factors.Those include medical conditions (e.G., diabetes, bruxism, etc.), patient habits (e.G., smoking, bad oral hygiene routine) and user error (e.G., surgical technique).There has not been any instance where either infection and/or bone loss, and when right conditions prevail and led to peri-implantitis whether reported or not along with the other two, have resulted from a manufacturing or design defect.Furthermore, the probability of manufacturing or design defects that might lead to infection and result in bone loss occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.The analysis and result of investigations and the analysis of probability previously described are contained in the summary investigation reports performed for bone loss and infection, which are attached.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.Dhr review was completed for the subject lot number 2020040358.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record op# 0210 was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 2020040358 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental : medical : peri-implantitis¿ as documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection, bone loss, and/or peri-implantitis problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report d9: device availability g3: date received by manufacturer g6: checked "follow-up" h2: checked follow-up type h3: changed "no" to "yes" h6: entered evaluation codes h10: added manufacturer narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT, 5MM(D) X 5MM(L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16768692
Report Number0001038806-2023-00778
Device Sequence Number19646420
Product Code DZE
UDI-Device Identifier00844868033639
UDI-Public(01)00844868033639(17)250825(10)2020040358
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K150571
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date08/25/2025
Device Model NumberBOES505
Device Catalogue NumberBOES505
Device Lot Number2020040358
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2023
Supplement Date Received by Manufacturer03/29/2023
Initial Report FDA Received Date04/19/2023
Supplement Report FDA Received Date09/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-