| Model Number |
27040GP1-S |
| Medical Device Problem Codes |
Material Separation (1562); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product has been discarded by the facility.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.The event is filed under internal karl storz complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
According to the information we received, the electrode item 27040gp1-s bipolar cutting loop was missing from the electrode.The loop was found inside the patient's bladder, it was retrieved and the turp procedure completed with no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Kst/mfr.Has provided the complaint product investigation plan and report which is based on the non return of product.Report conclusion root cause is according to the customer, the cutting loop was pulled out in the distal area.The reason for this could be, for example, excessive tensile force on the loop without having activated the hf current.However, since the article was not sent in for examination, no more precise statement can be made on this.The ifu points out the correct handling.The event is filed under internal karl storz complaint id (b)(4).
|
| |
|