Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Reporter is a synthes employee.Part: 03.161.003.Lot: l379230.Manufacturing site: werk hagendorf.Supplier: na.Release to warehouse date: 15 june 2017.Expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the occiput trial rod ø3.5 had signs of heavy scratches and discoloration of the blue anodize.The observed condition was consistent as an end of life indicator for the device.Additionally, the rod was broken between the third and fourth screw holes, both fragments were returned.The edges of the broken pieces present signs of repeated bending and reverse bending, these actions may have weakened the trial rod leading to breakage.After a visual inspection per guidance, it was determined that the reusable instrument device was scratched and discolored from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the occiput trial rod ø3.5 would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from colombia reports an event as follows: it was reported that several devices were received as blind units on (b)(6) 2023.Upon manufacturer investigation, it was determined that a trial rod had signs of heavy scratches and discoloration of the blue anodize.Additionally, he rod was broken between the third and fourth screw holes.This report is for a template for occipital plate/rod.This is report 2 of 2 for (b)(4).
|