| Model Number |
RNS-320-K - FGI |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The rns neurostimulator (model rns-320) reached end of life (eol) 3.3 years after initial implant (b)(6) 2019.Battery life expectancy is usually 6.0 - 12.4 years with typical use.With medium stimulation and detection settings the battery is estimated to last 10.8 years.The patient underwent a neurostimulator replacement on (b)(6) 2022.The device was returned and investigated by neuropace.When compared to expected battery plot behavior, the battery plot curve of the explanted device was observed to be discharging at an abnormally fast rate.The shape of the discharge curve was normal, except for downward spikes.
|
| |
|
Additional Manufacturer Narrative
|
|
The rns neurostimulator (model rns-320) reached end of life (eol) 3.3 years after initial implant (b)(6) 2019.Battery life expectancy is usually 6.0 - 12.4 years with typical use.With medium stimulation and detection settings the battery is estimated to last 10.8 years.The patient underwent a neurostimulator replacement on (b)(6) 2022.The device was returned and investigated by neuropace.When compared to expected battery plot behavior, the battery plot curve of the explanted device was observed to be discharging at an abnormally fast rate.The shape of the discharge curve was normal, except for downward spikes.
|
| |
|