Model Number CL-335-10-K
Medical Device Problem Codes Signal Artifact/Noise (1036); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
During an ecog review performed on 08/09/2022, an unusual signal was observed, indicative of a lead break on the right sub-temporal strip.The break appears to have occurred around feb 2022.There was no change in treatment, the doctor reported that the fracture did not seem to affect efficacy of the rns system.Troubleshooting included palpitation of the lead which resulted in high impedance and signal artifact.No known trauma was reported, and a cause was unable to be determined.The lead was secured with a burr hold cover.The lead has been implanted 1.7 years at time of lead break observation.On (b)(6) 2023 the patient underwent a lead revision.
 
Additional Manufacturer Narrative
(b)(4).The lead remains implanted but disconnected from the rns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Common Device Name
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
5108822607
MDR Report Key16769935
Report Number3004426659-2023-00015
Device Sequence Number6436889
Product Code PFN
UDI-Device Identifier00855547005304
UDI-Public010085554700530417230226
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCL-335-10-K
Device Catalogue Number1007934
Device Lot Number29716-1-1-1
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2023
Initial Report FDA Received Date04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age21 YR
Patient SexMale
-
-