Model Number CCWTT3
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was returned in the lens case.Solution was dried on the lens.Both haptics were broken/torn (possibly cut) in the haptic/optic junction area (returned) and in the distal tip area (not returned).The optic was torn/split/cracked and broken (possibly cut) into multiple pieces.Associated products were not provided.It is unknown if qualified products were used.The root cause could not be determined for the reported complaint of ¿defective¿.The specific lens issue was not specified.Damage to the lens was observed.While we are unable to determine the root cause of the lens damage, our observation reasonably suggests that it is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(6).
 
Event or Problem Description
A healthcare professional reported during an intraocular lens (iol) implant procedure, the lens was defective.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16770504
Report Number1119421-2023-00708
Device Sequence Number14266052
Product Code MJP
UDI-Device Identifier00380652457347
UDI-Public00380652457347
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/11/2026
Device Model NumberCCWTT3
Device Catalogue NumberCCWTT3.180
Device Lot Number15409687
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/23/2023
Supplement Date Received by Manufacturer04/11/2023
Initial Report FDA Received Date04/19/2023
Supplement Report FDA Received Date05/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/13/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-