| Model Number |
CCWTT3 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned in the lens case.Solution was dried on the lens.Both haptics were broken/torn (possibly cut) in the haptic/optic junction area (returned) and in the distal tip area (not returned).The optic was torn/split/cracked and broken (possibly cut) into multiple pieces.Associated products were not provided.It is unknown if qualified products were used.The root cause could not be determined for the reported complaint of ¿defective¿.The specific lens issue was not specified.Damage to the lens was observed.While we are unable to determine the root cause of the lens damage, our observation reasonably suggests that it is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(6).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported during an intraocular lens (iol) implant procedure, the lens was defective.Additional information was requested.
|
| |
|