| Model Number |
P10.7-50Q155 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that the nail successfully transported prior to the procedure, but the surgeon tried to recharge the nail in-situ and on the back table and the nail would not work.The patient was implanted with a new nail.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection of the returned nail revealed no visible external damage.X-ray of the internal components determined the anti-jam washer suffered damage.The damaged noted in the x-ray image likely occurred as a result over-retraction user error.A review of device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|