| Model Number |
4515-70-100 |
| Medical Device Problem Codes |
Break (1069); Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:((b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty surgical procedure, it was observed that while using the robotic-assisted solution base station, the foot pedal cord was "shorting out".It was working intermittently and erratically causing the software to move back and forth through screens.It was reported that this issue occurred during pointer acquisitions.It was reported that moving the cord when the foot pedal wasn't activating affected the operation.Holding the cord caused it to behave erratically.It was reported that the case went to manual instrumentation.There were a slight delay in the case in order to switch to manual instrumentation.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was evaluated and the reported condition was confirmed.The assignable root cause for the foot pedal cord shorting out is due to premature wear.The system is performing as intended following replacement of the foot pedal cord with a new one.
|
| |
|
Search Alerts/Recalls
|