Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994); Ambulation or Postural Difficulties (2544)
Date of Event 11/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial left hemiarthroplasty.Approximately 2 months post implantation, the patient was revised due to pain, difficulty ambulating, and loosening.During the revision, periprosthetic fracture to the tibia was noted.All components were replaced with conversion to total knee replacement without complications.
 
Additional Manufacturer Narrative
(b)(4).No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: continues with pain, ambulating with cane; x-ray: loosening of tibia component; tibial metaphysis fracture 1 cm below the bottom portion of the component.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARTIAL TIBIAL CEMENTED SIZE D LEFT MEDIAL
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16776368
Report Number0001825034-2023-00870
Device Sequence Number20517208
Product Code HSX
UDI-Device Identifier00880304812666
UDI-Public(01)00880304812666(17)310625(10)65066289
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K161592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number42538000401
Device Lot Number65066289
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2023
Initial Report FDA Received Date04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight58 KG
-
-