Model Number MA-15C
Medical Device Problem Codes Device-Device Incompatibility (2919); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unspecified surgical procedure it was observed that the bearing sleeve device could not be removed from the minimal access driver device.During in-house engineering evaluation it was observed that the device's part has physical damage and the device's temperature is more (heat).There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The bearing sleeve was evaluated and the reported condition that the device could not be removed from the driver device was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device's temperature was more (heat).The assignable root cause of this condition was determined to be traced to component failure due to wear.A review of the device history was performed and no non-conformances were detected related to the reported condition.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEARING SLEEVE, 15CM CURVED
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key16778055
Report Number1045834-2023-00597
Device Sequence Number6670459
Product Code HBE
UDI-Device Identifier00845384004325
UDI-Public00845384004325
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K042783
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMA-15C
Device Catalogue NumberMA-15C
Device Lot NumberK053113819
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2023
Initial Report FDA Received Date04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-