• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMRIS - DEERFIELD IMAGING, INC. IMRIS ORT200 OPERATING ROOM TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMRIS - DEERFIELD IMAGING, INC. IMRIS ORT200 OPERATING ROOM TABLE Back to Search Results
Model Number ORT200
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An imris service engineer assessed the operating room table at the customer site the following day (b)(6) 2023 and isolated the cause to an actuator component within the table's rotational locking mechanism.The service engineer removed and replaced the actuator component and functionally tested the table after the repair.The actuator was returned to the manufacturer.An internal corrective action has been initiated by the manufacturer to further address this issue.
 
Event or Problem Description
The customer reported that during a morning quality assurance / functional check, the operating room table's rotational locking mechanism did function properly to lock table position in place.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT200 OPERATING ROOM TABLE
Common Device Name
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC.
1230 chaska creek way
suite 100
chaska MN 55318
Manufacturer Contact
paul campbell
1230 chaska creek way
suite 100
chaska, MN 55318
7632036344
MDR Report Key16778977
Report Number3010326005-2023-00004
Device Sequence Number6820016
Product Code KXJ
UDI-Device Identifier00857534006349
UDI-Public(01)00857534006349
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberORT200
Device Catalogue Number114148-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2023
Initial Report FDA Received Date04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/25/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-