| Model Number |
ORT200 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An imris service engineer assessed the operating room table at the customer site the following day (b)(6) 2023 and isolated the cause to an actuator component within the table's rotational locking mechanism.The service engineer removed and replaced the actuator component and functionally tested the table after the repair.The actuator was returned to the manufacturer.An internal corrective action has been initiated by the manufacturer to further address this issue.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a morning quality assurance / functional check, the operating room table's rotational locking mechanism did function properly to lock table position in place.There was no patient involvement.
|
| |
|