• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WESTMED LLC WESTMED LLC; 1ML VENTED LUER SLIP SYRINGE, 25IU/ML BALANCED HEPARIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WESTMED LLC WESTMED LLC; 1ML VENTED LUER SLIP SYRINGE, 25IU/ML BALANCED HEPARIN Back to Search Results
Model Number 79720
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reportable, as this would contaminate the sample and cause a re-draw.
 
Event or Problem Description
The customer has noted multiple issues with the syringes: the syringe doesn't always fill properly.When aspirating, air gets into the sample (although they did state that this often happens with self-filling syringes).Blood also comes through the cap (blood dripped through the cap).Blood has sometimes also flowed through the plunger of the syringe.
 
Additional Manufacturer Narrative
Reportable, as this would contaminate the sample and cause a re-draw.Customer email indicated that the results of the analysis remains consistent with previous syringe.The customer was provided potential causes for partial syringe fill.Ra: rma-20036b risk id r37 indicates the risk associated with "partial fill of syringe due to user error" is a 7 - extreme.
 
Event or Problem Description
The customer has noted multiple issues with the syringes: 1.The syringe doesn't always fill properly 2.When aspirating, air gets into the sample (although they did state that this often happens with self-filling syringes).3.Blood also comes through the cap (blood dripped through the cap).4.Blood has sometimes also flowed through the plunger of the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WESTMED LLC
Common Device Name
1ML VENTED LUER SLIP SYRINGE, 25IU/ML BALANCED HEPARIN
Manufacturer (Section D)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer (Section G)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key16779818
Report Number2028807-2023-00010
Device Sequence Number6858446
Product Code CBT
UDI-Device Identifier00709078013287
UDI-Public00709078013287
Combination Product (Y/N)N
Initial Reporter CountryFI
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number79720
Device Catalogue Number11561340
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/10/2023
Supplement Date Received by Manufacturer04/10/2023
Initial Report FDA Received Date04/20/2023
Supplement Report FDA Received Date05/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-