| Model Number |
79720 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reportable, as this would contaminate the sample and cause a re-draw.
|
| |
|
Event or Problem Description
|
|
The customer has noted multiple issues with the syringes: the syringe doesn't always fill properly.When aspirating, air gets into the sample (although they did state that this often happens with self-filling syringes).Blood also comes through the cap (blood dripped through the cap).Blood has sometimes also flowed through the plunger of the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
Reportable, as this would contaminate the sample and cause a re-draw.Customer email indicated that the results of the analysis remains consistent with previous syringe.The customer was provided potential causes for partial syringe fill.Ra: rma-20036b risk id r37 indicates the risk associated with "partial fill of syringe due to user error" is a 7 - extreme.
|
| |
|
Event or Problem Description
|
|
The customer has noted multiple issues with the syringes: 1.The syringe doesn't always fill properly 2.When aspirating, air gets into the sample (although they did state that this often happens with self-filling syringes).3.Blood also comes through the cap (blood dripped through the cap).4.Blood has sometimes also flowed through the plunger of the syringe.
|
| |
|
Search Alerts/Recalls
|