• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL Back to Search Results
Model Number MCENDO550R
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H6: component code: 866 - lenses.Health effect - impact code: 4648 - insufficient information.Health effect - clinical code: 4580 - insufficient information.Medical device problem code: 3001 - optical problem.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
 
Event or Problem Description
The user facility reported to terumo cardiovascular a blurry endoscope.It is unknown when this event occurred, whether there was a delay in the procedure, whether the product was changed out, or if there was any effect on the patient results of the surgery.Due to the unknown information for this event, it is being reported.As per the user facility, no additional information will be provided.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on april 20, 2023.H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event or Problem Description
Blurry endoscope.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d3 (manufacturer - corrected email address) d4 (additional device information - added exp date) g1 (contact office - corrected first/given name, last name, email address, telephone number) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 ( device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 3331, 3224, 3259 ,19) type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 3331 - analysis of production records investigation findings #1: 3224 - optical problem identified investigation findings #2: 3259 - improper physical structure investigation conclusions: 19 - cause traced to user a visual observation of the internal components using a monocular was conducted on the returned sample.It was discovered that there was a crack on one of the internal lenses which resulted in a blurry visual field.Improper handling caused the internal lens to break.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURBISHED ENDOSCOPE 5.5M
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key16780399
Report Number1124841-2023-00093
Device Sequence Number7014261
Product Code GCJ
UDI-Device Identifier00699753023222
UDI-Public(01)00699753023222
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K092789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCENDO550R
Device Catalogue NumberN/A
Device Lot Number800373
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2023
Supplement Date Received by Manufacturer04/13/2023
05/25/2023
Initial Report FDA Received Date04/20/2023
Supplement Report FDA Received Date05/05/2023
05/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-